Undergraduatelibrary.org

The Effect of the Decontamination of Asthma Spacer Devices on their Function and their Suitability for Reuse in a Paediatric Emergency
Abstract
Objectives: The aims of this study were to determine the effects of dishwashing and
dishwashing frequency on the efficiency of spacer devices used with ipratropium inhalers by determining effects on emitted dose. Introduction: Recent evidence shows that the delivery of bronchodilators via a Metered
Dose Inhaler-spacer device combination is as effective as delivery via a nebuliser in paediatric asthma emergency cases, with certain added benefits. However uptake of this practice has been slow due in part to the perceived increased cost associated with use of the spacer devices. At present spacer devices are recommended for single patient use only and must either be given to the patient or discarded after use. A previous study indicated that sufficient decontamination of spacer devices could facilitate their reuse in emergency departments, thus providing significant cost savings. However, the effect of this decontamination procedure on spacer performance has to be determined before a change to practice could be implemented. Method: Three types of spacer devices were used in conjunction with ipratropium inhalers,
namely: Aerochambers®, Babyhalers® and Volumatics®. The emitted dose from an MDI- spacer device combination was measured before and after dishwashing. Emitted dose was measured via a Pharmacopeial method with drug detection being carried out via High Performance Liquid Chromatography analysis. Two wash cycles were undertaken with two separate sets of spacers: twice daily washing for five days and once daily washing for ten Results: In all cases the emitted dose decreased after dishwashing. This decrease in emitted
dose varied from 5-20% depending on spacer type and was statistically significant after ten washes compared to control (unwashed). Conclusion: Repeated dishwashing of spacers can lead to a small but statistically significant
decrease in the emitted dose delivered to paediatric patients. This would have to be considered if the practice was to be implemented into paediatric emergency departments. Word Count: 4901 Keywords: ipratropium bromide, spacer devices, paediatric emergency department, asthma, 1. Introduction
1.1 Asthma in Ireland Asthma is a chronic disease which is characterised by inflammation of the airways. This inflammation causes reversible obstruction of air flow making it difficult for the patient to breathe and causing them to suffer from breathlessness, coughing and wheezing (Busse 2007). In Ireland up to 20% of children are affected by asthma with 79% of these children suffering with uncontrolled asthma. This is reflected in the fact that children account for approximately 55% of the 20,000 cases of asthma that present to emergency departments every year (Asthma Society of Ireland 2013) (Asthma Society of Ireland 2011). Drugs used in the treatment of asthma are predominantly delivered by inhalation. 1.2 Drug Delivery in Asthma Over 70 million people worldwide use a metered dose inhaler (MDI) either alone or in combination with an inhaler add-on device (Terzano 2001). Inhaler add-on devices or spacer devices were developed to help patients achieve better coordination between breathing and inhaler actuation. Spacer devices allow younger patients to tidal breathe and decrease gastrointestinal deposition of the drug in children who are crying (Barry & O'Callaghan 2003, Barry et al. 1996, Brown et al. 1990, Hindle & Chrystyn 1994, Lavorini & Fontana 2009, Morgan et al. 1982, Warner et al. 1998, Wildhaber et al. 1999). Spacer devices also decrease the velocity of drug particles in turn allowing the evaporation of propellant thereby creating smaller drug particles which are less likely to impact on the back of the patient's throat and more likely to be delivered to the patient's lung (Dalby & Suman 2003). A Cochrane review has now shown definitively that the combination of an MDI and a spacer device is as effective as the use of wet nebulisation for the delivery of β agonists in children and adults suffering from mild to moderate exacerbations of asthma with nebulisers still being recommended for those patients with severe exacerbations (Cates et al. 2009, Mason et al. 2008) . In addition, the use of an MDI and spacer device is associated with a significantly shorter length of stay in the emergency department compared to the use of a nebuliser (Cates et al. 2009, Newman et al. 2002). Although spacer devices have many advantages as regards drug delivery as described above, there are complications and disadvantages to their use. Some of the problems commonly experienced with spacer devices are outlined below. Plastic and polycarbonate spacers are most widely used, however these can accrue a static charge on their walls due to their non-conducting nature (Piérart et al. 1999). This can cause charged drug particles to accumulate on the interior surface of the spacer device, which in turn decreases drug delivery to the lungs (Barry & O'Callaghan 2003, O'Callaghan et al. 1993). To avoid the build up of any static charge on the inside of the spacer device, the manufacturers of spacer devices recommend hand-washing the spacer devices with a mild detergent and allowing them to air dry (NHS Primary Care Trust 2011). An exception to this is the Aerochamber® which can be washed on the top rack of the dishwasher at temperatures not exceeding 70°C (Trundell Medical International 2012). The combination of an MDI and a spacer device is now considered to be of equal efficacy to wet nebulisation in the treatment of acute asthma exacerbations. Studies have shown a lack of uptake of this practise in many paediatric emergency departments (Dewar et al. 1999, Powell et al. 2001, Scott et al. 2009). While the reasons for this are manifold most studies cite cost as a major barrier to change. There is substantial conflict in the literature as regards actual cost comparisons of the two methods. Studies which have found that use of an MDI and spacer device is less costly than wet nebulisation are based mainly on the fact that use of this combination reduces admissions and time spent in the emergency department. This may be counteracted by the increased labour costs associated with the increased time required to deliver the drugs when using a spacer device (Bowton & Haponik 1992, Cates et al. 2009, Dhuper et al. 2011, Doan et al. 2011). 1.3 Reuse of Spacer Devices This study investigates whether spacer devices could be reused with a view to decreasing cost associated with their use. A study examining the effectiveness of various washing techniques on spacer decontamination found that with the correct washing technique cross contamination with re-use of spacer devices is unlikely to be a major issue (Blackburn et al. 2011). As well as this the risk of any microbiological contamination, or indeed cross- contamination with the spacer devices would be much less than that which would exist with the reuse of nebulisers (Scott et al. 2009). Anecdotal evidence suggests that at present spacer devices are given to the patient to be taken home with them as the manufacturers recommend that the devices are suitable for single patient use only. To date no other published studies have looked at the effects of dishwashing spacers on their performance, but anecdotal reports suggest that this has indeed been done before in some hospitals at no detriment to the patient concerned. Thus this study was undertaken to examine whether or not the spacer devices could be safely reused with a view to decreasing costs associated with their use. 2. Methods
2.1 Spacer Devices Three types of spacer devices were used in this study; the Babyhaler® and Volumatic® made by Allen and Hanbury's and Trundell Medical's Aerochamber Yellow®. Four of each spacer type were used giving a total of twelve spacer devices. Spacer devices were marked with a unique code using a permanent marker to ensure their identification throughout the study. 2.2 Emitted dose An emitted dose device connected to a flow controller and vacuum pump represents the Pharmacopeial standard for quality control testing of inhalers (Copley Scientific 2012, USP 32 2009) and thus this method was chosen. A vacuum flow rate of 28.3L/min was chosen as this rate is considered to be particularly appropriate for mirroring a child's breathing rate (Copley Scientific 2012, Dubus et al. 2001, Finlay 1998, Peyron et al. 2005, USP 32 2009). It was calculated that at a flow rate of 28.3L/min the vacuum pump would have to run for eight seconds to allow four litres of air to flow through the spacer device. Four litres mimics average inhalational capacity and is the amount recommended by the manufactures of the equipment (Copley Scientific 2012). Thus a time of eight seconds was set on the timer. As shown below in Fig. 1 the collection tube is a cylindrical apparatus. A glass fibre filter measuring 25mm in diameter and with a pore size of 1 micron was placed at one end of the collection tube. This filter prevented passage of the drug particles into the tubing attached to the flow controller and thus minimised loss of emitted drug in this way. A rubber mouthpiece adapter was attached to the other end of the collection tube which allowed for the attachment of the inhaler or spacer device to the collection tube. This setup is shown in Figures 1, 2, 3 and 4 below.


Two actuations of the inhaler were fired to waste outside the apparatus to prime the inhaler and ensure that it was working as recommended by the manufacturers (Boehringer Ingelheim Limited 2012). A thirty second interval was observed between actuations. This is representative of the time between actuations when the inhaler is being used in practice, allowing the patient to tidal breathe and this practise has been documented in the literature (Dubus et al. 2001). Eight actuations of the inhaler were used per run as this is the amount that would be used in practice for the treatment of an acute exacerbation of asthma in a child over six years of age in the emergency department (OLCHC 2012). When the run was complete, the vacuum pump was switched off and the inhaler and mouthpiece adapter removed. 10ml of Fisher Scientific HPLC Grade Water was poured into the collection tube. The collection tube was then sealed and inverted ten times to allow dissolution of the drug particles. Two 1ml samples were then taken from the collection tube using a sterile 1ml pipette. Each sample was filled in to a separate glass tube, which was then labelled. The tubes were stored in the fridge until High Performance Liquid Chromatography (HPLC) analysis could be undertaken. (Copley Scientific 2012) Fig. 1: Diagram showing the apparatus used for measurement of emitted dose.



Fig. 2: Experimental setup for the measurement of emitted dose when using an Aerochamber® spacer device. Fig. 3: Experimental setup used for measuring the emitted dose when using a Volumatic®


Fig. 4: Experimental setup used for measuring the emitted dose when using a Babyhaler® 2.3 Standard Solutions A standard of ipratropium bromide was ordered from LRC Standards to facilitate the making up of standard solutions, the choosing of a method of analysis and the creation of a standard curve. Standard solutions were made up from a stock solution. The concentrations of the standard solutions used were 7.5, 10, 15 and 25 µg/ml. 2.4 High Performance Liquid Chromatography 2.4.1 Buffer and Mobile Phase The HPLC method used was based on that outlined in the literature for the isolation and detection of ipratropium bromide and related compounds (Simms et al. 1998). Two mobile phases were used: Mobile Phase A and Mobile Phase B. These were made up from a common buffer which contained 100mM of KH2PO4 adjusted to pH 4 with orthophosphoric  1600ml buffer.  400ml HPLC Grade Acetonitrile.  550ml buffer.  450ml HPLC Grade Acetonitrile. The column used for elution was a Phenomenex Luna 5u C18 column of dimensions 100mm X 4.6mm. This column is similar to those used in the literature(Simms et al. 2.4.3 Running the HPLC The programme used for the elution of the ipratropium was as follows: An injection volume of 100μl was used as this provided a sufficient amount of drug for detection. The UV detector was set at 210nm as this is the optimal wavelength for detection as outlined in the literature (Brambilla et al. 1999, Majoral et al. 2007, Simms et al. 1998). Two injections from each sample were used. This allowed an average absorbance to be ascertained, thus increasing accuracy. A set of standards were run to determine the elution time of ipratropium bromide. The elution time was determined to be between 5.1 and 5.5 minutes. The standard solutions were then used to develop a standard curve. 2.4.4 Data Analysis At the end of each sample set the area under the peak at the relevant elution time was recorded. This area under the peak was then substituted into the equation obtained from the standard curve. Thus the micrograms of ipratropium per 1ml of sample were determined. As the collection tube had initially been washed out with 10ml of water the value obtained was multiplied by ten to find the total amount in the collection tube; the emitted dose. This value was then divided by 160μg, the maximum possible dose, and multiplied by 100 to get the percentage emitted dose. 2.5 Statistical Analysis The change in the emitted dose from each spacer device after washing was calculated. Four of each type of spacer device were used in this study. In each group of spacer devices an outlier was identified. This outlier was excluded from the final calculations. Thus all results are based on three devices of each spacer type. A one-tailed t-test was performed on Microsoft Excel and p values were ascertained. 95% confidence intervals were also calculated and these are included in the results. The dishwashing cycle was set at 60°C for forty five minutes. The spacers were then removed from the dishwasher and air dried. The detergent used was a normal household detergent. The valves on the Babyhaler® were not removed during washing as this would be unlikely to be done in a busy emergency department in reality. For one set of spacers the washing took place twice daily for five days. The other set of spacers were washed once daily for five days, retested and then washed once daily for a further five days, thus giving 10 days of once daily washing. 3. Results
3.1: Emitted dose of Atrovent® MDI The emitted dose of the Atrovent® MDI without a spacer device attached was ascertained using the method described above. This provided evidence that the label claim of the inhaler was being delivered. The emitted dose of each inhaler used in the study was recorded so as to ensure that any decrease in emitted dose seen after washing the spacer devices was not simply due to a decrease in the emitted dose of the inhaler. Table 1 below shows the average emitted dose for each of the Atrovent® MDI's used as well as the overall average and standard deviation. Table 1: Average emitted doses from Atrovent® MDI without spacer devices attached. Inhaler Number
% Emitted Dose
Standard Deviation 3.2 Effect of Once Daily Washing on Emitted Dose: In all cases % relates to the emitted dose obtained as a percentage of the maximum possible 3.2.1 Aerochambers®: The emitted dose of ipratropium was measured using the Atrovent® inhaler attached to the Yellow Aerochamber® device before the devices had been washed. Emitted dose was measured after five once daily washes and again after a further five once daily washes giving ten washes in total. The emitted doses obtained as well as the total decrease in emitted dose are shown in Table 2 below. Table 2: Change in emitted dose after once daily washing at five days and ten days. ** Indicates statistical significance with a p<0.01 when emitted dose after 10 once daily washes is compared with control. 3.2.2 Babyhalers®: Emitted dose of ipratropium was measured using the Atrovent® inhaler attached to the Babyhaler® device as described in the method. This testing was undertaken at the same time points as for the Aerochambers. The emitted doses obtained as well as the total decrease in emitted dose are shown in Table 3 below. Table 3: Change in emitted dose noted after once daily washing at five days and ten days. ** Indicates statistical significance with a p<0.01 when emitted dose after 10 washes is compared with control. 3.2.3 Volumatics®: Emitted dose of ipratropium was measured using the Atrovent® inhaler attached to the Volumatic® device as described in the method. Testing was undertaken before washing (control), after five once daily washes and again after a further five once daily washes giving ten washes in total. The emitted doses obtained as well as the total decrease in emitted dose are shown in Table 4 below. Table 4: Change in emitted dose noted after once daily washing at five days and ten days. ** represents a p<0.01 when emitted dose after 10 washes is compared with control. 3.2.4 Average change in emitted dose: Table 5 below shows a summary of the information contained in sections 3.2.1-3.2.3 above. P values are calculated with respect to the emitted dose obtained before any washing was Table 5: Average change in emitted dose and associated statistical significance compared to Change from
Spacer Type
Fig. 5: The average emitted dose of each of the three spacer types before washing, post five washes and post ten washes (n=3 +/- S.D.). Statistical significance is calculated with reference to the emitted dose before washing. **p<0.01. 3.3 Effect of Twice Daily Washing on Emitted Dose 3.3.1 Aerochambers®: Emitted dose of ipratropium was measured using the Atrovent® inhaler attached to the Yellow Aerochamber® device as described in the method. This testing was undertaken before washing (control) and after five days of twice daily washing. The emitted doses obtained as well as the total decrease in emitted dose are shown in Table 6 below. Table 6: Change in emitted dose and associated significance after 5 days of twice daily washing. * represents p<0.05 and is calculated with respect to the emitted dose obtained on control testing. Post 10 washes
Change from
3.3.2 Babyhalers®: Emitted dose of ipratropium was measured using the Atrovent® inhaler attached to the Babyhaler® device as outlined in the method section. Emitted dose was measured before washing (control) and after five days of twice daily washing. The emitted doses obtained as well as the total decrease in emitted dose are shown in Table 7 below. Table 7: Change in emitted dose and significance associated with said change after 5 days of twice daily washing. *represents p<0.05 and is calculated with reference to the emitted dose of the control. Post 10 washes
Change from
3.3.3 Volumatics®: Emitted dose of ipratropium was measured using the Atrovent® inhaler attached to the Volumatic® spacer device as outlined in the method section. Emitted dose was measured before washing (control) and after five days of twice daily washing. The emitted doses obtained as well as the total decrease in emitted dose are shown in Table 8 below. Table 8: Change in emitted dose and significance associated with said change after 5 days of twice daily washing. * shows a decrease in emitted dose that is statistically significant Post 10 washes
Change from
3.3.4 Average change in emitted dose after five days of twice daily washing: Table 9 gives an overview of the average drop off in emitted dose after five days of twice daily washing and the statistical significance associated with same. Fig. 6 shows this decrease graphically. Table 9 : Average change in emitted dose following five days of twice daily washing. Average change
Post 5 Days BD Washing (10 washes) Spacer type
Fig. 6: The emitted dose for each spacer device before washing and the emitted dose post ten washes. (n=3 +/- S.D.). p<0.05. 4. Discussion
The aim of this study was to determine the effect of dishwashing on spacer device effectiveness, by measuring the change in emitted dose from the spacer devices. The drug used in this study was ipratropium, marketed as the Atrovent® MDI. Ipratropium is a short acting atropine like bronchodilator which is used in patients who are experiencing a mild exacerbation of asthma where salbutamol alone is not effective or for children with moderate asthma exacerbations (Kasawar & Farooqui 2010). The combination of an MDI and spacer device is now recommended for the delivery of β agonists in the treatment of acute exacerbations of asthma (Cates et al. 2009, Powell et al. 2001, Raucci et al. 1993). The Aerochambers® used in this study are the only spacers licensed for dishwashing (Trundell Medical International 2012). Thus, for the other spacer devices the washing undertaken here was outside of the manufacturers specifications. However, this study would have been futile had the washing conditions not mirrored what would be done in practise. Additionally hand-washing alone was shown to be ineffective at decontaminating spacer devices to a sufficient standard to allow their reuse (Blackburn et al. 2011). There is a paucity of evidence in the published literature as regards the estimated emitted dose of ipratropium from the Atrovent® inhaler. Available studies suggest an average emitted dose of between 83 and 94% from the Atrovent® inhaler (Brambilla et al. 1999). The values obtained in this study are in agreement with this. Each time a new inhaler was used the emitted dose was tested to ensure consistency. Owing to the proximity of all results no account was taken of the slight variability in the emitted dose between inhalers. Although values for emitted dose from spacer devices vary greatly in the literature, in all cases the emitted dose from the combination of an MDI and spacer device is lower than that from the inhaler alone (Barry et al. 1996). As outlined above, two washing cycles were undertaken to establish whether the amount of washes per day would affect spacer performance. In the once daily cycle spacers were tested before washing, after five days of washing and again after a further five days of washing, thus giving ten washes in total. The reduction in emitted dose after five days of once daily washing was not statistically significant. However a statistically significant decrease in emitted dose did occur after a further five days of once daily washing (Fig5). This trend indicates that there is a connection between number of washes and emitted dose, with emitted dose decreasing with an increasing number of washes. A second set of spacers were washed twice daily for five days. Thus while the number of washes was the same as for those spacers which were washed once daily, the frequency of washing differed. The spacers were washed using the same washing conditions as those being washed once daily. The reduction in emitted dose for these spacers was, on average less than that seen at ten days for those spacers undergoing once daily washing. Of note, the decrease in emitted dose for the Volumatics® in both wash cycles was approximately 16.6%. This would indicate that while the frequency of washing is not important, a statistically significant drop off in emitted dose occurs on washing. All spacer devices that were washed twice daily had statistically significant decreases in emitted dose after ten washes. Although being washed more regularly did not adversely affect the function of these spacer devices it is likely that in practise twice daily washing of spacer devices would not occur as there would be neither the resources to facilitate this nor the usage to warrant Another, similar study was carried out in parallel to this study using salbutamol delivered via a Ventolin® inhaler and the same types of spacer devices as in this study. Twice daily washing of the spacers was undertaken for five days giving a total of ten washes. Washing of spacer devices took place in the same dishwasher as that used in this study, using the same washing cycle to allow direct comparison of the results. In the salbutamol study decreases in emitted dose of 11.7% for Aerochambers®, 9.8% for Babyhalers® and 10.6% for Volumatics® were observed (Keane et. al Unpublished). This shows that regardless of the drug being used there is a decrease in emitted dose after dishwashing the spacer devices. It has previously been noted in the literature that individual drug molecules may have an effect on the emitted dose and thus this may account for the variability in the diminution in emitted dose seen with the two drugs (Ahrens et al. 1995, Barry & O'Callaghan 2003). There is also an amount of intra- and inter spacer variability. This has also been noted in the literature with factors such as spacer volume, properties of the drug substance and properties of the inhaler device used being seen to play a key role in this variability (Ahrens et al. 1995, Barry et al. 1996, Barry & O'Callaghan 2003). The results of this study show that the washing of spacer devices in a dishwasher does cause a statistically significant decrease the emitted dose and that this diminution in dose depends on the number as opposed to the frequency of the washes. However in this case, statistics alone cannot be used to determine significance. In an emergency situation the dose administered will be determined both by the age of the child and by the severity of the symptoms being experienced. However if the patient does not respond then the dose will be increased until an improvement in the patient's condition is seen (British Thoracic Soceity 2012). Thus to determine the real significance of these results the spacers would have to be tested on patients to see how greatly, if at all the decrease in emitted dose affects their response to treatment. There are also some issues with regard to the practicality of washing and drying the spacer devices. All of the spacers are identical, thus each spacer would have to be marked with a unique code to ensure that it would not be washed more than the requisite number of times. A log of what spacer was washed when would be a solution to this issue. To prevent the build up of charge on the interior of the spacer device the spacers need to be air dried. Although air drying of the device was carried out in this study it is difficult to see how this could be done on a more constant basis as this air drying requires quite a lot of space. Therefore the practicality of washing the spacers will be restricted by the space available for drying and the human resources required to monitor the washing log. All of this would also have to be taken into account when determining the cost effectiveness of washing and reuse of the spacer devices. 5. Limitations
The method used for determination of the emitted dose in this experiment gives only quantitative data as regards the emitted dose and does not allow determination of where in the respiratory tract the drug will be deposited. Thus it may be argued that although the amount of drug emitted may not change significantly, the properties of the emitted drug In the study which looked at the decontamination of the spacer devices, spacers were hand washed before dishwashing. This pre-wash by hand was not carried out in this study as it would probably not be feasible in a busy hospital environment. This pre-wash would be unlikely to affect spacer performance but may affect decontamination. Thus the microbiological study may need to be repeated to confirm that dishwashing alone provides sufficient decontamination. The spacer devices in this study were washed only ten times and therefore their use beyond this point cannot be recommended on the basis of this study. 6. Conclusions/Future Work
The results of this study show that there is a decrease in the emitted dose from spacer devices post dishwashing. This decrease varies between 5-20%. Although the results obtained are statistically significant their clinical significance is not yet clear. However washing the spacer devices on more than ten occasions is likely to further decrease the emitted dose which may then be detrimental to the patient. Successful and seamless change from nebulisers to spacers can only be achieved if there is education of and agreement between all parties involved in this changeover (Dewar et al. 1999, Doherty et al. 2007, Powell et al. 2001, Scott et al. 2009). Staff should regularly be reminded of these guidelines and an audit undertaken to ensure that said protocol is being followed appropriately. Provided it is ascertained that dishwashing alone provides sufficient decontamination of the spacer devices proper guidelines should be drawn up as regards their reuse as shown in Spacers should be marked to ensure they are not washed on more than ten occasions. A rota should be drawn up and marked each time a particular spacer is washed. Calculation of the incremental cost effectiveness ratio should be undertaken to assess the cost effectiveness of the MDI and spacer combination. It should be noted that the guidelines relating to spacer use also apply to the treatment of asthma exacerbations in adults and thus such cost savings could also be executed in a hospital dealing predominantly with adult patients (Mason et al. 2008). Appendix 1
Flow Chart Depicting when Spacer Devices Should be Sent for Dishwashing. Spacer Device used Patient Sent Home Spacer Device will Spacer Device No Spacer Device or Longer Needed by Doesn't Have One Already Has One at patient on the ward Give Spacer Device Send Spacer Device Give Spacer Device Send Spacer Device Ahrens, R., Lux, C. & Bahl, T., 1995. Choosing the metered-dose inhaler spacer or holding chamber that matches the patient ' s need  : Evidence that the specific drug being delivered is an important consideration. Journal of allergy and clinical immunology, 96, Anon, 2009. U.S. Pharmacopeia, Asthma Society of Ireland, 2013. Asthma Clinic. Available at: Asthma Society of Ireland, 2011. Hse. Health Service Executive Asthma Information, Barry, P W & O'Callaghan, C., 2003. The influence of inhaler selection on efficacy of asthma therapies. Advanced drug delivery reviews, 55(7), pp.879–923. Available at: Barry, P W, Robertson, C.F. & O'Callaghan, C., 1993. Optimum use of a spacer device. Archives of disease in childhood, 69(6), pp.693–4. Available at: &rendertype=abstract. Barry, Peter W, Callaghan, C.O. & Royal, L., 1996. Inhalational drug delivery from seven different spacer devices. Thorax, 51, pp.835–840. Blackburn, C. et al., 2011. SAEM Abstracts, Plenary Session. Academic Emergency Medicine, 18, p.S84. Available at: http://doi.wiley.com/10.1111/j.1553-2712.2011.01073.x [Accessed January 24, 2013]. Boehringer Ingelheim Limited, 2012. Atrovent Summary of Product Characteristics. , pp.1–7. Bowton, L. & Haponik, F., 1992. Substitution of Metered Dose Inhalers for Hand Held Nebulizers. Chest, 101, pp.305–308. Brambilla, G. et al., 1999. Modulation of aerosol clouds produced by pressurised inhalation aerosols. International journal of pharmaceutics, 186(1), pp.53–61. Available at: British Thoracic Soceity, 2012. British Guidelines on the Management of Asthma, Brown, P.H. et al., 1990. Do large volume spacer devices reduce the systemic effects of high dose inhaled corticosteroids? Thorax, 45(10), pp.736–9. Available at: &rendertype=abstract. Busse, W., 2007. Expert Panel Report 3 (EPR-3): Guidelines for the Diagnosis and Management of Asthma–Summary Report 2007. Journal of Allergy and Clinical Immunology, 120(5), pp.S94–S138. Available at: Cates, C., Crilly, J. & Rowe, B., 2009. Holding chambers ( spacers ) versus nebulisers for beta- agonist treatment of acute asthma ( Review ). Cochrane Collaboration, (1). Copley Scientific, 2012. Quality Solutions for Who are Copley, Available at: http://www.copleyscientific.com/documents/ww/Inhaler Brochure 2012 (Low Dalby, R. & Suman, J., 2003. Inhalation therapy: technological milestones in asthma treatment. Advanced drug delivery reviews, 55(7), pp.779–91. Available at: Dewar, a L. et al., 1999a. A randomised controlled trial to assess the relative benefits of large volume spacers and nebulisers to treat acute asthma in hospital. Archives of disease in childhood, 80(5), pp.421–3. Available at: &rendertype=abstract. Dhuper, S. et al., 2011. Efficacy and cost comparisons of bronchodilatator administration between metered dose inhalers with disposable spacers and nebulizers for acute asthma treatment. The Journal of emergency medicine, 40(3), pp.247–55. Available at: http://www.ncbi.nlm.nih.gov/pubmed/19081697 [Accessed November 2, 2012]. Doan, Q., Shefrin, A. & Johnson, D., 2011. Cost-effectiveness of metered-dose inhalers for asthma exacerbations in the pediatric emergency department. Pediatrics, 127(5), pp.e1105–11. Available at: http://www.ncbi.nlm.nih.gov/pubmed/21464192 [Accessed November 20, 2012]. Doherty, S.R. et al., 2007. Evidence-based implementation of adult asthma guidelines in the emergency department: a controlled trial. Emergency medicine Australasia  : EMA, 19(1), pp.31–8. Available at: http://www.ncbi.nlm.nih.gov/pubmed/17305658 [Accessed December 4, 2012]. Dubus, J.C., Rhem, R. & Dolovich, M., 2001. Delivery of HFA and CFC salbutamol from spacer devices used in infancy. International journal of pharmaceutics, 222(1), pp.101–8. Available at: http://www.ncbi.nlm.nih.gov/pubmed/11404036. Finlay, W., 1998. Inertial sizing of aerosol inhaled during pediatric tidal breathing from an MDI with attached holding chamber. International Journal of Pharmaceutics, 168(2), pp.147–152. Available at: Hindle, M. & Chrystyn, H., 1994. Relative bioavailability of salbutamol to the lung following inhalation using metered dose inhalation methods and spacer devices. Thorax, 49(6), pp.549–53. Available at: &rendertype=abstract. Kasawar, G.B. & Farooqui, M., 2010. Development and validation of a stability indicating RP- HPLC method for the simultaneous determination of related substances of albuterol sulfate and ipratropium bromide in nasal solution. Journal of pharmaceutical and biomedical analysis, 52(1), pp.19–29. Available at: http://www.ncbi.nlm.nih.gov/pubmed/20045275 [Accessed January 23, 2013]. Keane, A et al, The Effect of the Decontamination of Asthma Spacer Devices on their Function with Salbutamol Sulphate Inhalers and their Suitability for Reuse in a Paediatric Emergency Department. Unpublished Work. Lavorini, F. & Fontana, G. a, 2009. Targeting drugs to the airways: The role of spacer devices. Expert opinion on drug delivery, 6(1), pp.91–102. Available at: Mason, N. et al., 2008. Nebulisers or spacers for the administration of bronchodilators to those with asthma attending emergency departments? Respiratory medicine, 102(7), pp.993–8. Available at: http://www.ncbi.nlm.nih.gov/pubmed/18396026 [Accessed December 4, 2012]. Morgan, M. et al., 1982. SHORT REPORTS Terbutaline aerosol given through pear spacer in acute severe asthma. British Medical Journal, 285(September), pp.849–850. Newman, K.B. et al., 2002. A comparison of albuterol administered by metered-dose inhaler and spacer with albuterol by nebulizer in adults presenting to an urban emergency department with acute asthma. Chest, 121(4), pp.1036–41. Available at: NHS Primary Care Trust, 2011. Patient Information Leaflet. , (Mdi). O'Callaghan, C. et al., 1993. Improvement in sodium cromoglycate delivery from a spacer device by use of an antistatic lining, immediate inhalation, and avoiding multiple actuations of drug. Thorax, 48(6), pp.603–6. Available at: &rendertype=abstract. Our Lady's Children's Hospital Crumlin, Asthma Treatment Flow Diagram, OLCHC, 2012. Peyron, I.D. et al., 2005. Development and performance of a new hydrofluoroalkane (HFA 134a)-based metered dose inhaler (MDI) of salmeterol. Respiratory medicine, 99 Suppl A, pp.S20–30. Available at: http://www.ncbi.nlm.nih.gov/pubmed/15777605 [Accessed December 4, 2012]. Piérart, F. et al., 1999. Washing plastic spacers in household detergent reduces electrostatic charge and greatly improves delivery. The European respiratory journal  : official journal of the European Society for Clinical Respiratory Physiology, 13(3), pp.673–8. Available at: http://www.ncbi.nlm.nih.gov/pubmed/10232445. Powell, C. V et al., 2001. Successful implementation of spacer treatment guideline for acute asthma. Archives of disease in childhood, 84(2), pp.142–6. Available at: &rendertype=abstract. Raucci, J.C. et al., 1993. Emergency Department Treatment of. Chest, 103, pp.665–672. Scott, S.D. et al., 2009. Barriers and supports to implementation of MDI/spacer use in nine Canadian pediatric emergency departments: a qualitative study. Implementation science  : IS, 4, p.65. Available at: &rendertype=abstract [Accessed December 4, 2012]. Simms, P.J. et al., 1998. The separation of ipratropium bromide and its related compounds. Journal of pharmaceutical and biomedical analysis, 17(4-5), pp.841–9. Available at: Terzano, C., 2001. Pressurized metered dose inhalers and add-on devices. Pulmonary pharmacology & therapeutics, 14(5), pp.351–66. Available at: http://www.ncbi.nlm.nih.gov/pubmed/11603949 [Accessed December 2, 2012]. Trundell Medical International, 2012. AeroChamber Plus Flow-Vu Patieant Information Warner, J.O., Naspitz, C.K. & Cropp, G.J. a., 1998. Third International Pediatric Consensus statement on the management of childhood asthma. Pediatric Pulmonology, 25(1), pp.1–17. Available at: http://doi.wiley.com/10.1002/%28SICI%291099- Wildhaber, J.H. et al., 1999. Inhalation therapy in asthma: nebulizer or pressurized metered- dose inhaler with holding chamber? In vivo comparison of lung deposition in children. The Journal of pediatrics, 135(1), pp.28–33. Available at:

Source: http://www.undergraduatelibrary.org/system/files/2925_0.pdf

axa-tel.ba

Electronic Modular Switching System Please read this manual before connecting theElectronic Modular Switching System. ELECTRONIC MODULAR SWITCHING SYSTEM Thank you for purchasing the Panasonic Model KX-T206E, Electronic Modular Switching System. Electronic Modular Switching System Proprietary telephone with display

Microsoft word - publikationsliste-für-www-2007.doc

Prof.Dr.med. Stephan H. Duda Herausgeberschaften Gastherausgeberschaft im Themenheft „Primäre Lebertumoren" in „Der Onkologe", zusammen mit P.M. Schlag (2000) Editorial Board Mitglied "Investigative Radiology" (seit Januar 2002) Editorial Board Mitglied "Techniques in Vascular and Interventional Radiology" (seit Januar 2005) 1. Publikationen a.) Wissenschaftliche Originalarbeiten in referierten Journals 1. Biewald, W., S.H. Duda: Die Meatoplastik. Ein Konzept zur Behandlung der kongenitalen distalen