Heartstartmrxtds_hospital_rev5.fm
No compromises. That's how we created the
It is our mission to produce the finest quality
HeartStart MRx, a combination transport
medical equipment, enabling you to deliver
monitor, cardiograph, manual defibrillator, and
effective care and ultimately save lives. We
AED. HeartStart MRx unites our industry-
strive to meet the evolving needs of the
leading patient monitoring capabilities with
healthcare community with outstanding
our patented biphasic defibrillation therapy in
technology and innovation, while always
one comprehensive, easy-to-use, rugged, and
focusing on what's most important - positive
lightweight device, optimized to meet a
patient outcomes. So, whether it's the long-
variety of user needs.
life batteries, automated operational checks,
or a data management solution, HeartStart
With carefully arranged controls, connections,
MRx has the capabilities you need and the
and menus, HeartStart MRx efficiently
performance you demand - no compromises.
provides comprehensive diagnostic
measurements without compromising critical,
lifesaving therapy. Monitoring and
resuscitation functions are clearly delineated,
with monitoring located on one side of the
device and defibrillation, pacing, and
cardioversion on the other. Patient status and
care data --- including numeric vital signs and
up to four waves, as well as text prompts,
alarms, an event timer, event markers, and
battery status indicators --- are logically
organized on the device's large, backlit, color
Instructions for use (User's guide)
Quick reference cards
User training workbook
ST/AR Basic algorithm for arrhythmia detection
Application notes
ECG monitoring through monitoring electrodes and defibrillation pads
AC power module
Synchronized cardioversion
Adjustable ECG size and autogain
Manual and AED operation
HeartStart MRx Monitor/Defibrillator
SMART Biphasic waveform for defibrillation therapy
Ordering Option Information
Large 4-wave color display
Strip chart printer
Individual, adjustable volume of QRS beeper, voice prompts, and alerts
SpO2, NBP, and EtCO2
Configuration mode
12-Lead ECG Acquisition
Standard External Paddles
Operational checks
Water Resistant External Paddles
Automated self-tests with "ready-for-use" indicator
Lithium ion battery with capacity gauge
Additional Battery
ECG out capability
Barrel style Pad Cable (Replacement for Standard Pad
Accessory Pouch and Strap
SpO2 with Fourier Artifact Suppression Technology (FAST)
Noninvasive Blood Pressure
Instructions for Use
Microstream™ EtCO
User Training Video (English only)
Alternate Label Set, AED/Energy Select
Noninvasive Pacing
Service Manual (English only)
12-Lead ECG with Philips interpretation algorithm
Service Training Video (English only)
Lithium ion battery with capacity gauge
Hands-free multifunction electrode cable
3-Lead ECG cable
Disposable monitoring electrodes
Strip chart recorder paper
Defibrillator test load
Multifunction Electrode Pads
Adult Radiotransparent/Reduced Skin
Pediatric Radiotransparent/Reduced Skin
Pads Cable
Defib Hands-free, plug style
ECG Monitoring Electrodes
High-Tack Foam; 5 electrodes/pack (60 packs/case)
Option B02 - Acquisition
Option B03 - Transmission
ECG Cables
Option B04 - Wider Printer
12-Lead Cable Set
External Paddles with Paddle Tray
10-lead ECG Trunk Cable, 12-pin Connector (for 3-Lead
Option C01 - Standard Paddles
Option C02 - Water Resistant Paddles
3-lead ECG Set and Plug with Snap (AAMI)
Add 7-lead ECG Set for 12-Lead use (AAMI)
3-lead ECG Set and Plug with Snap (IEC)
Standard External Paddles
Add 7-lead ECG Set for 12-Lead use (IEC)
Water Resistant External Paddles
10-lead ECG Patient Trunk Cable, 12-pin ECG Input
Sterilizable External Paddles
Connector (for 5-Lead and 12-Lead use)
10-electrode Cable Set, Extremities, Grabber (use with
7.5 cm Switchless
10-electrode Cable Set, Chest, Grabber (use with
6.0 cm Switchless
4.5 cm Switchless
10-electrode Cable Set, Extremities, Grabber (use with
2.8 cm Switchless
10-electrode Cable Set, Chest, Grabber (use with
3-Lead Cable Set
3-lead ECG Trunk Cable (AAMI)
Large Disposable Switched
3-lead ECG Lead Set with Snaps (AAMI)
Medium Disposable Switched
3-lead ECG Trunk Cable (IEC)
Small Disposable Switched
3-lead ECG Lead Set with Snaps (IEC)
Large Disposable Switchless
5-Lead Cable Set
Medium Disposable Switchless
5-lead ECG Trunk Cable (AAMI)
Small Disposable Switchless
5-lead ECG Lead Set with Snaps (AAMI)
Internal Paddles Adapter Cable
5-lead ECG Trunk Cable (IEC)
5-lead ECG Lead Set with Snaps (IEC)
Multifunction Electrode Pads
Adult Radiolucent
Sync Cables
Multi-Patient Comfort Cuffs - Pediatric
Sync Cable (8 ft.)
Multi-Patient Comfort Cuffs - Small Adult
Sync Cable (25 ft.)
Multi-Patient Comfort Cuffs - Adult
Multi-Patient Comfort Cuffs - Large Adult
Reusable SpO2 Sensor - Adult Finger
Disposable Blood Pressure Cuffs
Reusable SpO2 Sensor - Pediatric/Small Adult Finger
Soft Single-Patient Disposable Cuff - Infant
Reusable SpO2 Sensor - Adult Thumb
Soft Single-Patient Disposable Cuff - Pediatric
Soft Single-Patient Disposable Cuff - Small Adult
2 Sensor - Adult Ear Clip
Soft Single-Patient Disposable Cuff - Adult
2 Sensor - Infant
Soft Single-Patient Disposable Cuff - Adult X-Long
Soft Single-Patient Disposable Cuff - Large Adult
SpO2 Extension Cable, 2 m
Soft Single-Patient Disposable Cuff - Large Adult X-
Nellcor SpO2 Sensor Adapter Cable, 1 m (use with
Soft Single-Patient Disposable Cuff - Thigh
Adult Pressure (5 ft.)
Filter Line Set - Adult/Pediatric (25 sets/case)
Adult Pressure (10 ft.)
Filter H Set - Adult/Pediatric (25 sets/case)
Filter H Set - Infant/Neonatal (yellow, 25 sets/case)
Reusable Blood Pressure Cuffs
Reusable NBP Cuff Kit, 3 sizes (pediatric, adult, large
Non-Intubated Dual Purpose Circuits (CO2 + O2)
Reusable NBP Cuff Kit, 5 sizes
Smart CapnoLine - Pediatric
(infant, pediatric, adult, large adult, thigh)
Smart CapnoLine - Intermediate
Traditional Reusable NBP Cuff - Infant
Smart CapnoLine - Adult
Traditional Reusable NBP Cuff - Pediatric
Traditional Reusable NBP Cuff - Adult
Non-Intubated Single Purpose Circuits (CO2)
Traditional Reusable NBP Cuff - Large Adult
Smart CapnoLine - Pediatric
Traditional Reusable NBP Cuff - Thigh
Smart CapnoLine - Intermediate
Smart CapnoLine - Adult
Antimicrobial Reusable NBP Cuff - Infant
Antimicrobial Reusable NBP Cuff - Pediatric
Antimicrobial Reusable NBP Cuff - Small Adult
Lithium Ion Battery with Capacity Gauge
Antimicrobial Reusable NBP Cuff - Adult
Antimicrobial Reusable NBP Cuff - Large Adult
Antimicrobial Reusable NBP Cuff - Thigh
50 mm Chemical Thermal, Gray Grid (10 rolls)
50 mm Chemical Thermal, Gray Grid (80 rolls)
Color Handle
Green and color label set
Blue and color label set
Yellow and color label set
Rose and color label set
Gray and color label set
Test Load for use with M3508A Pad Cable
Carrying Case (includes 3 accessory pouches and
Data Card and Tray
Charge Time: Less than 5 seconds to 200 Joules with a new, fully charged
Lithium Ion battery pack at 25ºC. Less than 15 seconds when operating
without a battery, using the M3539A AC power module alone at 90% rated
Dimensions with pads: 12.4 in. (W) x 8.3 in. (D) x 11.7 in. (H)
Patient Impedance Range:
(31.5 cm x 21.0 cm x 29.5 cm)
Minimum: 15 Ohm (internal defibrillation); 25 Ohm (external defibrillation)
Dimensions with paddles: 13.4 in. (W) x 8.3 in. (D) x 13.6 in. (H)
(34.0 cm x 21.0 cm x 34.5 cm)
Weight: 13.2 lbs. including pads, pads cable, full roll of paper, and battery.
Note: Actual functional range may exceed the above values.
Incremental weight of external standard paddles and paddle tray is 2.5 lbs.
Additional battery weighs 1.6 lbs.
Manual Mode
Manual Output Energy (Selected): 1-10, 15, 20, 30, 50, 70, 100, 120,
150, 170, 200J; maximum energy limited to 50J for internal defibrillation
Waveform: Biphasic Truncated Exponential. Waveform parameters adjusted
as a function of patient impedance
Controls: On/Off Therapy Knob, Charge, Shock, Sync, Print, Mark Event, ECG
Lead Select, Alarm Pause, Event Review
Shock Delivery: Via multifunction defib electrode pads, or paddles
Energy Selection Control: Front panel therapy knob
Delivered Energy Accuracy:
Charge Control: Front panel button; button on external paddles
Nominal Delivered Energy vs. Patient Impedance
Shock Control: Front panel button; buttons on external or switched internal
Load Impedance
Synchronized Control: Front panel SYNC button
100 125 150 175
Indicators: Text prompts, audio alerts, QRS beeper, battery status, Ready For
Use, external power/battery charging, Sync mode
Armed Indicators: Charging tone, charged tone, flashing shock button, and
energy level indicated on display
AED Energy Profile: 150 Joules nominal into a 50 Ohm test load
Text and Voice Prompts: Extensive text/audible messages guide user
through configured protocol
AED Controls: On/Off, Shock
Indicators:
Monitor display messages and prompts, voice prompts, battery
status, Ready For Use, external power/battery charging
Armed Indicators: Flashing shock button, charged tone, and voice prompts
ECG Analysis: Evaluates patient ECG and signal quality to determine if a
shock is appropriate and evaluates connection impedance for proper
defibrillation pad contact
Shockable Rhythms: Ventricular fibrillation and certain ventricular
tachycardias, including ventricular flutter and polymorphic ventricular
Shock Advisory Algorithm Sensitivity and Specificity: Meets AAMI DF-
ECG and Arrhythmia Monitoring
Inputs: Up to four (4) ECG waves may be viewed on display and up to 2
Size: 8.4 in. diagonal (128 mm x 171 mm)
waves printed simultaneously. Lead I, II, or III is obtained through the 3-lead
Type: TFT Color LCD
ECG cable. With a 5-lead cable, leads aVR, aVL, aVF, and V can also be
obtained. Pads ECG is obtained through 2 multifunction defibrillation electrode
Resolution: 480 x 640 pixels (VGA)
Sweep Speed: 25 mm/s nominal (stationary trace; sweeping
Lead Fault: Leads Off message and dashed line appear on the display if an
erase bar) for ECG and SpO2; 6.25 mm/s for CO2
electrode or lead wire becomes disconnected. Lead Off indicator in wave
Wave Viewing Time: 5 seconds (ECG)
Pad Fault: Dashed line appears on the display if a pad becomes
Type: 6.3 Ah, 14.8 V, rechargeable, Lithium Ion
Heart Rate Display: Digital readout on display from 15 to 300 bpm, with an
accuracy of ±10%
Dimensions: 6.5 in. (H) x 3.8 in. (W) x 1.6 in. (D) (165 mm x 95 mm x 42
Heart Rate/Arrhythmia Alarms: HR, Asystole, VFIB/VTACH, VTACH,
Extreme Tachy, Extreme Brady, PVC rate, Pacer not capture, Pacer not pacing
Weight: 1.6 lb. (0.73 kg)
Hands Free Defibrillation Cable Length:
Charge Time: Approximately 3 hours to 100%. Approximately
2 hours to 80%, indicated by battery gauge indicator. Charging the battery at
M3508A = 5 ft. (1.6 m)
temperatures above 45ºC may degrade battery life.
M3507A = 7 ft. (2.2 m)
Capacity: At least 5 hours of monitoring with ECG, SpO2, and CO2 monitored
ECG Cable Length: 9 ft. (2.7 m)
continuously and NBP measured every 15 minutes, or at least 50 200-joule
Common Mode Rejection: Greater than 90 dB measured per AAMI
shocks (on a new, fully charged battery at room temperature, 25º C). At least
standard for cardiac monitors (EC 13)
3.5 hours while pacing at 180 ppm at 160 mA and monitoring as described
ECG Size: 2.5, 5, 10, 20, 40 mm/mV, autogain
Battery Indicators: Battery gauge on battery, capacity indicator on display;
Frequency Response:
flashing RFU indicator, chirp, and LOW BATTERY message appears on display
AC Line Filter: 60 Hz or 50 Hz
for low battery condition*
Pads ECG for Display: Monitor (0.15-40 Hz) or EMS (1-30 Hz)
*Low battery condition triggered when 10 minutes of monitoring time and 6
Pads ECG for Printer: Monitor (0.15-40 Hz) or EMS (1-30 Hz)
maximum energy discharges remain (with a new battery at room temperature,
Leads ECG for Display: Monitor (0.15-40 Hz) or EMS (1-30 Hz)
Leads ECG for Printer: Diagnostic (0.05-150 Hz) or Monitor (0.15-40 Hz) or
Storage: Storing the battery for extended periods at temperatures above
40ºC will reduce battery capacity and degrade battery life.
Patient Isolation (defibrillation proof):
External Defib: Type BF
Internal Defib: Type CF
Thermal Array Printer
Noninvasive Blood Pressure
Continuous ECG Strip: The Print key starts and stops the strip. The printer
can be configured to run real time or with a 10-second delay. The strip prints
Pressure Range:
the primary ECG lead with event annotations and measurements.
Systolic: 40-260 mmHg
Auto Printing: The printer can be configured to automatically print on
Diastolic: 20-200 mmHg
Marked Events, Charge, Shock, and Alarm. When an alarm condition occurs,
Initial Pressure: 160 mmHg (Adult); 120 mmHg (Pediatric)
the unit prints the primary ECG wave and the alarming wave, if configured.
Maximum Pressure: 280 mmHg
Overpressure Safety Limits: Maximum of 300 mmHg
Reports: The following can be printed: Event Summary, 12-Lead, Operational
Cuff Inflation Time: 50 second maximum (pediatric or adult)
Check, Configuration, Status Log, and Device Information
Accuracy: ±3 mmHg
Speed: 25 or 50 mm/s with an accuracy of ± 5%
Alarm Range:
Systolic high limit: 35 - 270 (Adult), 35 - 180 (Pediatric)
Amplitude Accuracy: ± 5% or ± 40 uV, whichever is greater
Systolic low limit: 30 - 265 (Adult), 30 - 175 (Pediatric)
Paper Size: 50 mm (W) by 30 m (100 ft.) (L)
Diastolic high limit: 15 - 245 (Adult), 15 - 150 (Pediatric)
Diastolic low limit: 10 - 240 (Adult), 10 - 145 (Pediatric)
Mean high limit: 25 - 255 (Adult), 25 - 160 (Pediatric)
Mean low limit: 20 - 250 (Adult), 20 - 155 (Pediatric)
Waveform: Monophasic Truncated Exponential
Rated Life: 50,000 measurement cycles (36/day for 2.3 years)
Current Pulse Amplitude: 10 mA to 160 mA (5 mA resolution);
Auto Mode Repetition Time: 1, 2.5, 5, 10, 15, 30, 60, or 120 minutes
accuracy 10% or 5 mA, whichever is greater
Measurement Time: Auto/manual mode: 30 seconds (average) @ HR > 60
bpm, 170 seconds (maximum)
Pulse Width: 40 ms with ± 10% accuracy
Interconnect Tube Length:
Rate: 30 ppm to 180 ppm (10 ppm increments); accuracy ± 1.5%
M1598B Connect tubing (1.5 m)
Modes: Demand or Fixed Rate
M1599B Connect tubing (3 m)
Refractory Period: 340 msec (30 to 80 ppm); 240 msec (90 to 180 ppm)
End-Tidal Carbon Dioxide
SpO2 Pulse Oximetry
Range: 0 to 99 mmHg
Resolution: 1mmHg (0.1kPa)
Range: SpO2: 0-100%; pulse rate: 30 to 300 bpm
Accuracy: For values between 0 and 38 mmHg: ±2 mmHg. For values
between 39 and 99 mmHg: ±5% of reading + 0.08% for every 1 mmHg
Display Update Period: 1 sec. typical numeric update rate
(above 40 mmHg). Values read at sea level after > 20 min. warm up
Alarm Range:
Accuracy with:
Low Limit: 10 to 94 mmHg (Adult/Pediatric)
M1191A sensor - 1 standard deviation 70% to 100%, ± 2.5%
High Limit: 20 to 95 mmHg (Adult/Pediatric)
M1192A sensor - 1 standard deviation 70% to 100%, ± 2.5%
Sample Size: 50 ml per min.
M1193A sensor - 1 standard deviation 70% to 100%, ± 2.5%
M1194A sensor - 1 standard deviation 70% to 100%, ± 4.0%
M1195A sensor - 1 standard deviation 70% to 100%, ± 4.0%
NELLCOR sensors - 1 standard deviation 80% to 100%, ± 3.0%
Pulse Rate Accuracy: 2% or 1 bpm (whichever is greater)
Alarm Range:
Low Limit: 50 to 99% (Adult/Pediatric)
High Limit: 51 to 100% (Adult/Pediatric)
Alarm Delay: 10 seconds
Airway Respiration Rate
Range: 0 to 150 rpm
Temperature: 0º C to 45ºC operating, -20º to 70ºC storage
Resolution: 1 rpm
Humidity: Up to 95% relative humidity
Accuracy:
0 to 40 rpm ±1 rpm
Altitude:
41 to 70 rpm ±2 rpm
Operating: 0 to 15,000 ft. (0 to 4,500 m)
71 to 100rpm ± 3%
Storage: 0 to 15,000 ft. (0 to 4,500 m)
101 to 150rpm ± 5%
Shock:
Operating: Half-sine waveform, duration < 3 ms, velocity change > 111 in/s
Alarm Range:
±5% 3 times on all six faces
Low Limit: 0 to 99 rpm (Adult/Pediatric)
Storage: Trapezoidal waveform, acceleration > 30 g, velocity change=742 cm/
High Limit: 10 to 100 rpm (Adult/Pediatric)
s ±10% on all six faces
Apnea Alarm Time: 10-40 seconds, in increments of 5
Vibration:
Operating: Random vibration, 0.30 grms, 5-500 Hz for >10 minutes/axis
Calibration Gas for CO2 Measurement System
PSD=0.0002 g2/Hz from 5 to 350 Hz, -6 dB/octave slope from 350 to 500 Hz
Ingredients: 5% Carbon Dioxide, 21% Oxygen, 74% Nitrogen
Storage: Random vibration, 2.41 grms, 5-500 Hz for >10 minutes/axis
Cylinder Size: BD
PSD=0.02 g2/Hz from 5 to 100 Hz, -6 dB/octave slope from 100 to 137 Hz,
Method of Preparation: Gravimetric
0 dB/octave slope from 137 to 350 Hz, -6 dB/octave slope from 350 to 500
Blend Tolerance: 0.03%
Accuracy: 0.03% absolute
Moisture: 10 PPM Maximum
Solids Resistance: IP2x
Expiration Period: 2 years
Water Resistance: Meets EN 60601-2-4. Water testing performed with
Pressure: 144 PSIG, Volume: 1L
patient cables connected to the device
EMC: Meets EN 60601-1-2:2001
12-Lead ECG
Safety: Meets EN 60601-1, UL 2601-1, CAN/CSA C22.2 No. 601-1
Inputs: With a 10-Lead cable, leads I, II, III, aVR, aVL, aVF, V/C1-V/C6 can be
obtained. All 12-Lead ECG waves can be viewed on the display simultaneously.
Other Considerations: Device not suitable for use in the presence of
All 12 leads can be printed on the strip chart printer in 3x4 format.
concentrated oxygen or a flammable anesthetic mixture with air, oxygen, or
12-Lead ECG Transmission: Data card to a PC
Receiver Type: HeartStart Event Review Pro
Mode of Operation: Continuous
Patient Data Storage
AC Line Powered:
Input: 100-240 VAC, 50-60 Hz, 1-0.46 A
Internal Event Summary: 8 hours of continuous ECG waveforms and
Output: 18V, 5A, 90W
events AND 50 12-Lead ECG reports
Battery Powered: 14.8 V Rechargeable, Lithium Ion
Data Card Event Summary: 8 hours of continuous ECG waveforms and
events AND 50 12-Lead ECG reports
Philips Medical Systems is part
Philips Medical Systems
of Royal Philips Electronics
Cardiac and Monitoring Systems3000 Minuteman RoadAndover, MA 01810Tel: (800) 934-7372
Philips Medical Systems2660 Matheson Blvd. E.
Mississauga, Ontario L4W 5M2Tel: (800) 291-6743
EUROPE, MIDDLE EAST AND AFRICA:
Philips Medizinsysteme Böblingen GmbHCardiac and Monitoring SystemsHewlett-Packard Str. 271034 BöblingenGermanyFax: (+49) 7031 463 1552
LATIN AMERICA HEADQUARTERS:
Philips Medical Systems5200 Blue Lagoon Drive9th FloorMiami, FL 33126Tel: (305) 267-4220
ASIA PACIFIC HEADQUARTERS:
Philips Medical Systems26/F Hopewell Centre17 Kennedy RoadWanchaiHong KongTel: (+852) 2821 5888Fax: (+852) 2527 7136
2003 Koninklijke Philips Electronics North America. All rights are reserved. Reproduction in whole or in part is prohibited without the prior written consent of the copyright holder. Koninklijke Philips Electronics N.V. reserves the right to make changes in specifications or to discontinue any product at any time without notice or obligation and will not be liable for any consequences resulting from the use of this publication.
Published December 20034522 981 84631
Source: http://www.laerdal.fi/binaries/AASOOKBT.pdf
UP .Universidad de Palermo Facultad de Ciencias Sociales Trabajo final integrador Percepción de riesgo y prevalencia de consumo de drogas en alumnos universitarios Tutor: Dr. Eduardo Leiderman Autor: Andrés Otero 3-2-1-Consumo de sustancias en subregión de América del sur………….…………… 9 3-2-2-El consumo de sustancias de los universitarios argentinos……….…………. 10
North American Tour Presented in association with UCLA Film & Television Archive and supported in part by grants from the Getty Foundation and The Andy Warhol Foundation for the Visual Arts. The series is curated by Al yson Nadia Field, Jan-Christopher Horak, Shannon Kelley, and Jacqueline Stewart. The Andy Warhol Foundation for the Visual Arts L.A. RebeLLioN Comes Home